Skip Navigation

This Article
Right arrow Abstract Freely available
Right arrow FREE Full Text (PDF) Freely available
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrow Search for citing articles in:
ISI Web of Science (7)
Right arrowRequest Permissions
Google Scholar
Right arrow Articles by Pierce, K. E.
Right arrow Articles by Wangh, L. J.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Pierce, K. E.
Right arrow Articles by Wangh, L. J.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

Molecular Human Reproduction, Vol. 9, No. 12, pp. 815-820, 2003
© European Society of Human Reproduction and Embryology 2003; all rights reserved

Detection of cystic fibrosis alleles from single cells using molecular beacons and a novel method of asymmetric real-time PCR

Kenneth E. Pierce1, John E. Rice, J.Aquiles Sanchez and Lawrence J. Wangh

Department of Biology, MS-008, Brandeis University, Waltham, MA 02454-9110, USA

1 To whom correspondence should be addressed. e-mail: pierce{at}brandeis.edu


    Abstract
 Top
 Abstract
 Introduction
 Materials and methods
 Results
 Discussion
 References
 
We present a method for rapid and accurate identification of the normal and {Delta}F508 alleles of the cystic fibrosis (CF) gene in single human cells that utilizes LATE (linear after the exponential)-PCR, a newly invented form of asymmetric PCR. Detection of the single-stranded amplicon is carried out in real time, using allele-specific molecular beacons. The LATE-PCR method permits controlled abrupt transition from exponential to linear amplification and thereby enhances the fluorescent signals and reduces variability between replicate samples relative to those obtained using typical real-time PCR. Of 239 single lymphoblasts generating amplification signals, 227 (95%) exhibited signals that met objective quantitative criteria required for diagnosis. Among these samples, 222 were genotyped correctly, for an assay accuracy of 98%. The small number of diagnostic errors was due to allele drop-out among heterozygous lymphoblasts, 4/119 (3.4%), and contamination among homozygous {Delta}F508 lymphoblasts, 1/57 (1.8%). LATE-PCR offers a new strategy for preimplantation genetic diagnosis and other fields in which accurate quantitative detection of single copy genes is important.

Key words: cystic fibrosis {Delta}F508/fluorescent oligonucleotide probes/linear DNA amplification/preimplantation genetic diagnosis/quantitative PCR


    Introduction
 Top
 Abstract
 Introduction
 Materials and methods
 Results
 Discussion
 References
 
Cystic fibrosis (CF) is the most common inherited disease among Caucasian populations with an incidence of ~1 in 2500 births (Welsh et al., 1995). Couples in which both individuals carry a mutant copy of the CF gene have a one in four chance of having an affected child. The conditions caused by these mutations range from mild to life-threatening. Preimplantation genetic diagnosis (PGD) offers such couples an opportunity to determine the genetic composition of their embryos before starting a pregnancy and thereby avoid either having an affected child or terminating a pregnancy of an affected fetus. However, PGD is technically difficult because each assay is carried out on a single cell recovered from a cleavage stage embryo.

A 3 base pair (bp) deletion, designated {Delta}F508, accounts for nearly 70% of CF cases and causes severe manifestations of the disease. It results in the absence of phenylalanine at position 508 of the cystic fibrosis transmembrane conductance regulator (CFTR) protein and this error prevents normal processing and translocation of the polypeptide chain to apical membranes of epithelial cells (Cheng et al., 1990). The first tests for {Delta}F508 in single cells used nested PCR to amplify the requisite sequence followed by verification of the final product either by restriction enzyme digestion (Coutelle et al., 1989), hybridization to allele-specific oligonucleotides (Wu et al., 1993), or by heteroduplex formation (Liu et al., 1993; Avner et al., 1994). The first clinical reports of PGD for CF also utilized heteroduplex analysis of the PCR products (Handyside et al., 1992; Verlinsky et al., 1992; Ao et al., 1996). More recent PCR assays have used fluorescently labelled primers to increase sensitivity and reduce the rate of allele drop-out (ADO), a failure to amplify one allele from a heterozygous cell (Findlay et al., 1995; Verlinsky and Kuliev, 2000; Goossens et al., 2000). When this approach is employed, the fluorescently labelled products are separated and identified by electrophoresis after PCR amplification is finished.

In this report, we describe the use of real-time PCR with molecular beacons to identify the normal and {Delta}F508 alleles of cystic fibrosis in single human cells. Molecular beacons are fluorescently tagged single-stranded oligonucleotides that only fluoresce when bound to a complementary sequence. Several molecular beacons, each with a differently coloured fluorophore, can be used simultaneously to monitor accumulation of PCR products that differ by as little as a single base (Tyagi and Kramer, 1996; Kostrikis et al., 1998) and thereby offer exquisite allele-discrimination capabilities. Molecular beacon assays are carried out in closed tubes that permit fluorescence detection during product amplification, thereby shortening the duration of the assay and minimizing the risk of laboratory contamination.

In order to increase the intensity of the molecular beacon signals and reduce sample variation, we developed a unique PCR method in which amplification of a double-stranded DNA molecule abruptly shifts to linear amplification of the single strand to which the molecular beacons bind. We call this new method LATE-PCR, an acronym for ‘linear after the exponential’. As with all asymmetric PCR, LATE-PCR uses primers at different concentrations, but unlike typical asymmetric PCR, amplification is efficient due to improved primer design. LATE-PCR with molecular beacons also has similarities to symmetric real-time PCR assays, such as those developed for detection of the human Y chromosome (Pierce et al., 2000) and detection of Tay–Sachs disease (Rice et al., 2002), in that both generate signals that reflect the number of target molecules present in the initial sample. However, LATE-PCR consistently generates strong signals because the absence of product strand reannealing permits unhindered hybridization of the molecular beacon to its target strand and continued accumulation of that strand beyond the cycle at which symmetric reactions typically plateau (Sanchez et al., 2003).


    Materials and methods
 Top
 Abstract
 Introduction
 Materials and methods
 Results
 Discussion
 References
 
Lymphoblasts
Human lymphoblast cell lines homozygous or heterozygous for CF {Delta}F508, or homozygous for the normal CF gene sequence, were obtained from Coriell Cell Repositories (USA). Lymphoblasts were prepared and manipulated as described elsewhere (Pierce et al., 2000). Single lymphoblasts were transferred directly to MicroAmp optical PCR tubes (PE Applied Biosystems, USA) containing 10 µl QuantiLyse (Hamilton Thorne Biosciences, USA), an optimized lysis solution containing proteinase K (Pierce et al., 2002). Samples were incubated at 50°C for 30 min, then 95°C for 10 min in an ABI 7700 (PE Applied Biosystems).

Molecular beacons and primers
Molecular beacons were designed according to previously described methods (Tyagi and Kramer, 1996; Pierce et al., 2000) and were purchased from Biosearch Technologies (Novato, USA). The molecular beacon sequence for the normal allele was 5'-FAM-CGCGCTTATCATCTTTGGTGTTT CCTATAGCGCG-Dabcyl-3' and for the {Delta}F508 allele was 5'-TET-CG CGCTAAAATATCATTGGTGTTTCCTAAGCGCG-Dabcyl-3'. FAM and TET are fluorescent molecules and Dabcyl quenches that fluorescence when the beacon is in a hairpin configuration in the absence of its target sequence. Primers were designed to amplify an 84 bp region that includes the {Delta}F508 location. The upper primer sequence was 5'-CCTGGATTATGCCTGG CACCAT-3' and the lower primer sequence was 5'-CCTTGATG ACGCTTCTGTATCTA-3'. The melting temperatures of these primers were estimated using a formula based on nearest-neighbour thermodynamics (Allawi and SantaLucia, 1997) and a salt concentration adjustment (SantaLucia et al., 1996). A program that computes these values is available at http://bioweb.pasteur.fr/seqanal/interfaces/melting.html. The Tm of the upper primer at the limiting concentration of 50 nmol/l is 59°C and the Tm of the lower primer at a concentration of 1 µmol/l is 58°C, each calculated at a total monovalent cation concentration of 70 mmol/l. Specific amplification of the 84 bp product from single lymphoblasts was confirmed using SYBR Green detection during amplification and product denaturation, and by electrophoresis through agarose gels.

PCR conditions
Fifteen microlitres of concentrated PCR reagent mixture was added to each tube containing a lysed cell (or no cell control) to yield a final sample volume of 25 µl with final concentrations of 1xPCR buffer (Invitrogen, USA), 3.75 mmol/l MgCl2, 0.25 mmol/l dATP, 0.25 mmol/l dCTP, 0.25 mmol/l dGTP, 0.75 mmol/l dUTP, 50 nmol/l upper primer, 1 µmol/l lower primer, 1.2 µmol/l of each molecular beacon, and 1.5 IU Platinum Taq DNA polymerase (Invitrogen). Amplification and fluorescence detection were carried out in an ABI 7700. Thermal cycling consisted of an initial 5 min denaturation at 95°C followed by four cycles of 95°C for 10 s, 55°C for 2 min, and 72°C for 30 s, followed by 21 cycles of 95°C for 10 s, 55°C for 30 s, and 72°C for 30 s, followed by 35 cycles of 95°C for 10 s, 52°C for 30 s, and 72°C for 30 s, with fluorescence acquisition during the 52°C step. The total duration of the PCR is 3 h. Contamination control procedures for reagent preparation and sample handling were described previously (Pierce et al., 2000).


    Results
 Top
 Abstract
 Introduction
 Materials and methods
 Results
 Discussion
 References
 
Previous reports from our laboratory have emphasized that real-time PCR is particularly useful for PGD because kinetic analysis allows optimization of PCR conditions otherwise not possible (Pierce et al., 2000; Rice et al., 2002). Efficient amplification is indicated by early detection signals with minimal variance among replicates. The present analysis of cystic fibrosis began with an evaluation of several pairs of primers at equimolar concentration and several molecular beacons under a variety of concentrations and buffer conditions. None of these proved entirely satisfactory because signal strength was low. In order to overcome this problem we explored the use of asymmetric PCR in which primers are used at unequal concentrations in order to generate a single-stranded product that is more accessible to a molecular beacon.

We realized that reducing the concentration of an oligonucleotide reduces its melting temperature (Tm), a fact that is not taken into account in traditional asymmetric PCR. We discovered that the primer used at limiting concentrations must have a concentration-adjusted Tm equal to, or above, that of the primer used in excess in order to achieve efficient exponential amplification of both DNA strands and then abruptly switch to linear amplification when the limiting primer is exhausted. Moreover, by adjusting the concentration of the limiting primer, it is possible to cause this switch to occur coincident with the cycle at which the fluorescent signals reach the detection threshold, i.e. the CT value. Accumulation of the single-stranded product can be monitored by including one or more molecular beacons that target this strand without competition from the complementary strand. This protocol, designated LATE-PCR, is an improvement over traditional approaches to asymmetric PCR. LATE-PCR is particularly useful for analysis of single target molecules because it generates brighter molecular beacon signals, reduces sample variance, and allows amplification to continue for many more cycles without plateau (Sanchez et al., 2003).

All reactions described here contained two molecular beacons. One, labelled with the fluorophore TET, was complementary to 22 nucleotides at the site of the {Delta}F508 3 bp deletion; the other, labelled with the fluorophore FAM, was complementary to 22 nucleotides in the same region of the normal allele. The specificity of each molecular beacon for single target sequences was confirmed by PCR analysis of single lymphoblasts homozygous for either the normal allele, or the {Delta}F508 allele. As anticipated, homozygous normal lymphoblasts generated fluorescent signals from only the molecular beacon specific to the normal allele, while fluorescence from the molecular beacon to the {Delta}F508 allele always remained below threshold (green lines in Figure 1A, B). Conversely, lymphoblasts homozygous for {Delta}F508 generated fluorescent signal from the molecular beacon specific to the {Delta}F508 allele, while fluorescence from the normal allele beacon remained below threshold in the vast majority of samples (red lines in Figure 1A, B). Figure 1 also illustrates that the increasing signal observed with LATE-PCR is linear rather than sigmoidal.



View larger version (43K):
[in this window]
[in a new window]
 
Figure 1. Examples of specific molecular beacon fluorescence increase during real-time PCR in samples containing single lymphoblasts homozygous normal for CF (green), heterozygous {Delta}F508 (blue), or homozygous {Delta}F508 (red). (A) Fluorescent signal from the molecular beacon detecting the normal allele. (B) Fluorescent signal from the molecular beacon detecting the {Delta}F508 allele. Dashed lines indicate the threshold of 200 units (~10 SD above baseline readings) used for determining CT values.

 
When lymphoblasts heterozygous for the {Delta}F508 and normal alleles of cystic fibrosis were assayed by LATE-PCR, all but a few samples generated both signals (blue lines in Figure 1A, B). Comparison of the three sets of signals in Figure 1 reveals that the CT values of most reactions from heterozygous cells (blue lines) are slightly higher than the CT values of reactions from cells homozygous for either allele (green and red lines). In addition, the mean slope of the lines from heterozygous cell reactions is approximately one-half of the mean slope of the lines from homozygous cells. These observations are consistent with the fact that heterozygous cells contain only a single copy of each allele whereas homozygous cells contain two copies of one of the alleles. The mean strength of the signals from heterozygous cells is more than double that obtained using symmetric PCR and the SD is vastly reduced (Pierce et al., 2001).

The above results illustrate the quantitative dimensions that real-time LATE-PCR technology adds to PCR analysis. The kinetics of signal detection provide additional information that is useful for sample analysis. Since clinical samples with unknown genotypes can only be evaluated based on the signals generated (normal only, mutant only, or both), each lymphoblast sample was analysed on that basis, regardless of its known genotype. Although most reactions within each of these groups had similar kinetics, a few samples generated CT and/or final fluorescence values far from the mean values for that set. Such variants may reflect problems with target availability, PCR amplification, or detection of products. In the context of PGD, such deviants could increase the chances of a misdiagnosis. Therefore, each set of reactions was evaluated in terms of both CT values and final fluorescence (fluorescence intensity after 60 PCR cycles). The resulting analysis established objective quantitative criteria with which to judge whether a particular reaction should or should not be considered diagnosable. Samples having either a CT value or a final fluorescence value >3 SD from the mean values for the corresponding set were categorized as atypical and were not used to evaluate the accuracy of the LATE-PCR assay (see below).

Figure 2 displays the final fluorescence values of all lymphoblast samples, as well as the boundaries 3 SD away from each mean value. The set of reactions generating only normal allele signals (box 1) includes all homozygous normal lymphoblasts (green diamonds) plus four of the 125 heterozygous lymphoblasts (blue circles). The latter are examples of ADO, as they did not generate the expected {Delta}F508 signal. In addition, one lymphoblast homozygous for {Delta}F508 (red triangles) unexpectedly generated only a normal allele signal, presumably the result of contamination combined with failure to either transfer the lymphoblast to the PCR tube, or to amplify the {Delta}F508 allele within that cell. The CT values and final fluorescence for these normal allele signals are presented in Figure 3A. The contaminated sample, one of the heterozygous samples, and two homozygous normal samples exhibited CT values >3 SD from the mean CT value and were eliminated from further evaluation based on this objective criterion.



View larger version (18K):
[in this window]
[in a new window]
 
Figure 2. Final fluorescence values from normal and {Delta}F508 allele-specific molecular beacons following amplification from single lymphoblasts homozygous normal for the CF gene (green diamonds), heterozygous {Delta}F508 (blue circles), or homozygous {Delta}F508 (red triangles). Dashed lines indicate 3 SD from the means of samples generating only normal signal (box 1), only {Delta}F508 signal (box 2), or both signals (box 3).

 


View larger version (20K):
[in this window]
[in a new window]
 
Figure 3. Scatter plots of CT values and final fluorescence from single lymphoblast samples generating (A) only normal allele, or (B) only {Delta}F508 allele signal. Lymphoblasts include those homozygous normal for the CF gene (open diamonds), heterozygous {Delta}F508 (filled circles), or homozygous {Delta}F508 (open triangles). Dashed lines indicate 3 SD from the means.

 
The final fluorescence values for all samples exhibiting only the {Delta}F508 allele signal are presented in box 2 of Figure 2. These data include the great majority of the homozygous {Delta}F508 lymphoblasts (red triangles), as well as one heterozygous lymphoblast (blue circle) that failed to generate a signal for the normal allele, a case of ADO. All final fluorescence and CT values for all points in this data set are within 3 SD of the corresponding means (Figure 3B).

Also included in Figure 2 are the final fluorescence values for all reactions generating both signals. These data include 120 heterozygous lymphoblasts (blue circles), of which 117 had values within the threshold boundaries (box 3). Three samples displayed preferential amplification of the {Delta}F508 allele and therefore fell above the upper boundary of the box. Four homozygous {Delta}F508 lymphoblasts also generated both signals, presumably due to contamination. However, for these samples the final fluorescence values for the normal allele were among the lowest observed and for {Delta}F508 were of unusually high intensity in two of the four samples, placing them above the upper boundary of box 3. All samples that fell within box 3 were evaluated further by examining CT values for both signals (Figure 4). Two heterozygous lymphoblasts (filled circles) and one of the two remaining homozygous {Delta}F508 lymphoblasts (open triangles) had a CT value >3 SD above the mean, as indicated by the symbols outside the box. Thus, only one of the four samples contaminated with normal DNA could not be excluded as atypical based on either final fluorescence or CT values. Of the 120 heterozygous samples, 115 yielded both signals with final fluorescence and CT values within 3 SD of the means.



View larger version (15K):
[in this window]
[in a new window]
 
Figure 4. CT values for normal and {Delta}F508 allele signals from single lymphoblast samples that are within the 3 SD limits for final fluorescence of both signals (i.e. are within box 3 in Figure 2). This data set includes 117 heterozygous {Delta}F508 lymphoblasts (filled circles) and two homozygous {Delta}F508 lymphoblasts (open triangles). Dashed lines indicate 3 SD from the means.

 
These results are summarized in Table I. Of the total 256 lymphoblasts tested, 239 exhibited at least one signal above threshold, yielding an assay utility of 93%. The utility ranged from 83 to 98% depending on the cell line used, suggesting that the PCR assay per se was not responsible for the majority of the amplification failures. Of the 239 samples that generated at least one fluorescent signal, 227 (95%) had CT and final fluorescence values within 3 SD of the appropriate mean values. We use the term diagnostic efficiency to refer to the fraction of samples generating signals within these limits and recommend that only such samples should be used for clinical diagnostic purposes (Pierce et al., 2000). Had the 227 diagnosable samples presented clinically, four heterozygotes would have been misdiagnosed as homozygotes due to ADO, and one {Delta}F508 homozygote would have been misdiagnosed as heterozygous due to contamination. Thus, the LATE-PCR assay described here has a diagnostic accuracy of 98% (222/227).


View this table:
[in this window]
[in a new window]
 
Table I. Evaluation of real-time PCR for diagnosis of cystic fibrosis (CF) genotype
 

    Discussion
 Top
 Abstract
 Introduction
 Materials and methods
 Results
 Discussion
 References
 
Molecular beacons are effective tools for identifying individual alleles in real-time closed-tube reactions (Kostrikis et al., 1998). The results presented here demonstrate that the normal and {Delta}F508 alleles of the cystic fibrosis gene in single cells can be distinguished rapidly and accurately using this technology in combination with LATE-PCR. Improved accuracy is achieved, in part, by utilizing the quantitative features of real-time detection. Samples that do not exhibit amplification kinetics within a defined range can be objectively excluded from diagnosis. We previously used such criteria to develop an assay that was >99% accurate for detection of the multi-copy TSPY gene on the human Y chromosome in lymphocytes, as well as in blastomeres from embryos having <30% fragmentation (Pierce et al., 2000). We have now extended the same analytic approach to alleles of a single copy gene in single cells. Quantitative analysis of the data allowed us to objectively eliminate samples with atypical signals and thereby improve the accuracy of the assay from 96 to 98%. Although a few samples that gave the expected fluorescent signals were also eliminated from diagnostic evaluation, we believe that the increase in accuracy is worth the 5% reduction in diagnosable samples, even if a small percentage of normal embryos is excluded from transfer when this assay is applied to PGD. Similar improvement in accuracy was obtained when this strategy was used for the most common mutant allele of the Tay–Sachs gene (Rice et al., 2002).

The ADO rate observed among the 119 heterozygous lymphoblasts with diagnosable signals was 3.4% (4/119). This rate is lower than rates observed using conventional PCR and similar to the best rates observed using fluorescent primers (Lissens and Sermon, 1997; Goossens et al., 2000; Verlinsky and Kuliev, 2000). One reason for low ADO is likely to be the use of QuantiLyse, a proteinase K-based lysis solution that has been optimized using real-time PCR to ensure maximum availability of DNA targets for amplification (Pierce et al., 2002). Cell lysis in water as well as some versions of solutions containing proteinase K that have been used for PGD were clearly inferior for releasing DNA for amplification. Alkaline lysis solutions containing dithiothreitol (DTT) also gave inferior results, possibly due to the effect of residual DTT on PCR efficiency. Another possible reason for our low ADO rate rests in optimization of PCR conditions using real-time technology (Pierce et al., 2000), since the choice of thermal cycling profile and amplicon size are among factors that influence ADO rates (Piyamongkol et al., 2003). In addition, the kinetics of real-time amplifications can be used to evaluate the validity of the signals generated. For instance, one of the heterozygous {Delta}F508 lymphoblasts exhibited a single, late signal that could be distinguished from that of known homozygotes. In the case of Tay–Sachs heterozygotes, one of two examples of ADO could be similarly identified based on atypical kinetics (Rice et al., 2002). Conventional PCR strategies only analyse final PCR products without visualizing the kinetics of the reaction and therefore inherently have higher risk of misdiagnosis.

Another advantage of real-time PCR is the reduced risk of laboratory contamination with amplicon molecules, because the tubes are not opened following amplification. Consistent with this decreased risk of amplicon contamination, no signals were observed in any of 60 no-cell control reactions run after a large number of amplification reactions had already been carried out in our laboratory during the development of the CF assay.

Contamination of assays with cellular DNA is another possible serious problem in PGD. Five of 64 reactions prepared with homozygous {Delta}F508 lymphoblasts showed normal allele signals. However, by analysing the kinetics of amplification in these reactions, all but one was found to have at least one atypical signal, illustrating yet another advantage of the real-time PCR assay. Thus, the contamination rate in the data set presented here was 1.8% of the homozygous {Delta}F508 cells that had diagnosable signals, or 0.4% of all diagnosable samples. We anticipate lower contamination rates in the future as we focus on improving protocols for handling large numbers of single cell samples.

This study describes the first application of a generalizable design for real-time amplifications, which we have named LATE-PCR. The method results in linear accumulation of single-stranded DNA, and when used in conjunction with molecular beacons, increased signal intensity and reduced sample variation. These features are particularly useful for real-time PCR initiated with single cells. Standard real-time PCR requires ~40 cycles to reach levels of detection using molecular beacons, but typical exponential amplification begins to slow down and plateau within a small number of additional cycles (Rice et al., 2002). LATE-PCR circumvents this amplification plateau, enabling continued increase in signal for many additional cycles while minimizing the increased sample-to-sample variation typically observed at the end of PCR. LATE-PCR is related to asymmetric PCR (Gyllensten and Erlich, 1988) in the use of primers at different concentrations. However, traditional asymmetric PCR is most often done using primers designed for symmetric PCR. Reducing the concentration of one of those primers reduces amplification efficiency, a generally accepted phenomenon that can be verified using real-time detection (unpublished data). LATE-PCR design includes the use of a limiting primer that has a concentration-adjusted Tm at least as high as that of the excess primer. This enables efficient exponential amplification until the limiting primer is depleted and the reaction makes an abrupt transition from exponential to linear amplification. By adjusting the concentration of the limiting primer, this transition can be timed to coincide with the threshold cycle at which the target strand is first detected. Additional LATE-PCR design features are published elsewhere (Sanchez et al., 2003).

The new technologies described here promise higher accuracy, reduced handling, and reduced total assay time for clinical PGD. These improvements can be achieved using the most recent generation of lower cost, real-time PCR equipment (unpublished data), suggesting that PGD could soon be a service available at more IVF clinics. In order to make this goal a reality, assays for additional alleles of CF, Tay–Sachs and other genetic diseases will have to be developed. We are currently developing such assays using the LATE-PCR strategy because of its improved sensitivity and low ADO rates.


    Acknowledgements
 
The authors thank Dr Fred Kramer and Dr Sanjay Tyagi for assistance with molecular beacon technology. This research was supported in part by a grant from Hamilton Thorne Biosciences, Inc.


    References
 Top
 Abstract
 Introduction
 Materials and methods
 Results
 Discussion
 References
 
Allawi, H.T. and SantaLucia, J. (1997) Thermodynamics and NMR of internal G·T mismatches in DNA. Biochemistry, 36, 10581–10594.[CrossRef][Medline]

Ao, A., Ray, P., Harper, J., Lesko, J., Paraschos, T., Atkinson, G., Soussis, I., Taylor, D., Handyside, A., Hughes, M. and Winston, R.M.L. (1996) Clinical experience with preimplantation genetic diagnosis of cystic fibrosis ({Delta}F508). Prenat. Diagn., 16, 137–142.[CrossRef][ISI][Medline]

Avner, R., Laufer, N., Safran, A., Kerem, B-S., Friedmann, A. and Mitrani-Rosenbaum, S. (1994) Preimplantation diagnosis of cystic fibrosis by simultaneous detection of the W1282X and {Delta}F508 mutations. Hum. Reprod., 9, 1676–1680.[Abstract/Free Full Text]

Cheng, S.H., Gregory, R.J., Marshall, J., Sucharita, P., Souza, D.W., White, G.A., O’Riordan, C.R. and Smith, A.E. (1990) Defective intracellular transport and processing of CFTR is the molecular basis of most cystic fibrosis. Cell, 63, 827–834.[CrossRef][ISI][Medline]

Coutelle, C., Williams, C., Handyside, A., Hardy, K., Winston, R. and Williamson, R. (1989) Genetic analysis of DNA from single human oocytes: a model for preimplantation diagnosis of cystic fibrosis. Br. Med. J., 299, 22–24.[ISI][Medline]

Findlay, I., Ray, P., Quirke, P., Rutherford, A. and Lilford, R. (1995) Allelic drop-out and preferential amplification in single cells and human blastomeres: implications for preimplantation diagnosis of sex and cystic fibrosis. Mol. Hum. Reprod., 1, 1609–1618.

Goossens, V., Sermon, K., Lissens, W., Vandervorst, M., Vanderfaeillie, A., De Rijcke, M., De Vos, A., Henderix, P., Van de Velde, H., Van Steirteghem, A. and Liebaers, I. (2000) Clinical application of preimplantation genetic diagnosis for cystic fibrosis. Prenat. Diagn., 20, 571–581.[CrossRef][ISI][Medline]

Gyllensten, U.B. and Erlich, H.A. (1988) Generation of single-stranded DNA by the polymerase chain reaction and its application to direct sequencing of the HLA-DQA locus. Proc. Natl Acad. Sci. USA, 85, 7652–7656.[Abstract/Free Full Text]

Handyside, A.H., Lesko, J.G., Tarin, J.J., Winston, R.M.L. and Hughes, MR. (1992) Birth of a normal girl after in vitro fertilization and preimplantation diagnostic testing for cystic fibrosis. N. Engl. J. Med., 327, 905–909.[Abstract]

Kostrikis, L.G., Tyagi, S., Mhlanga, M.M., Ho, D.D. and Kramer, F.R. (1998) Spectral genotyping of human alleles. Science, 279, 1228–1229.[Free Full Text]

Lissens, W. and Sermon, K. (1997) Preimplantation genetic diagnosis: current status and new developments. Hum. Reprod., 12, 1756–1761.[Abstract/Free Full Text]

Liu, J., Lissens, W., Devroey, P., Van Steirteghem, A. and Liebaers, I. (1993) Polymerase chain reaction analysis of the cystic fibrosis {Delta}F508 mutation in human blastomeres following oocyte injection of a single sperm from a carrier. Prenat. Diagn., 13, 873–880.[ISI][Medline]

Pierce, K.E., Rice, J.E., Sanchez, J.A., Brenner, C. and Wangh, L.W. (2000) Real-time PCR using molecular beacons for accurate detection of the Y chromosome in single human blastomeres. Mol. Hum. Reprod., 6, 1155–1164.[Abstract/Free Full Text]

Pierce, K.E., Rice, J.E., Sanchez, J.A. and Wangh, L.W. (2001) A novel detection strategy for preimplantation genetic diagnosis (PGD) of cystic fibrosis and other diseases in single human cells. Fertil. Steril., 76 (Suppl. 1), S136.

Pierce, K.E., Rice, J.E., Sanchez, J.A. and Wangh, L.W. (2002) QuantiLyseTM: reliable DNA amplification from single cells. Biotechniques, 32, 1106–1111.[ISI][Medline]

Piyamongkol, W., Bermudez, M.G., Harper, J.C. and Wells, D. (2003) Detailed investigation of factors influencing amplification efficiency and allele drop-out in single cell PCR: implications for preimplantation genetic diagnosis. Mol. Hum. Reprod., 9, 411–420.[Abstract/Free Full Text]

Rice, J.E., Sanchez, J.A., Pierce, K.E. and Wangh, L.W. (2002) Real-time PCR with molecular beacons provides a highly accurate assay for Tay–Sachs alleles in single cells. Prenat. Diagn., 22, 1130–1134.[CrossRef][ISI][Medline]

Sanchez, J.A., Pierce, K.E., Rice, J.E. and Wangh, L.W. (2003) LATE PCR: a new strategy for enhancing the sensitivity and reliability of single-gene allele detection from single cells. Submitted for publication.

SantaLucia, J., Allawi, H.T. and Seneviratne, P.A. (1996) Improved nearest-neighbor parameters for predicting DNA duplex stability. Biochemistry, 35, 3555–3562.[CrossRef][Medline]

Tyagi, S. and Kramer, F.R. (1996) Molecular beacons: probes that fluoresce upon hybridization. Nature Biotechol., 14, 303–308.[CrossRef][ISI][Medline]

Verlinsky, Y. and Kuliev, A. (2000) An Atlas of Preimplantation Genetics. Parthenon Publishing, New York, p. 46.

Verlinsky, Y., Rechitsky, S., Evsikov, S., White, M., Cielak, J., Lifchez, A., Valle, J., Moise, J. and Strom, C.M. (1992) Preconception and preimplantation diagnosis for cystic fibrosis. Prenat. Diagn., 12, 103–110.[ISI][Medline]

Welsh, M.J., Tsui, L.-P., Boat, T.F. and Beaudet, A.L. (1995) Cystic fibrosis. In Scriver, C.R. et al. (eds), The Metabolic and Molecular Basis of Inherited Disease. McGraw-Hill, New York, vol. III, pp. 3799–3876.

Wu, R., Cuppens, H., Buyse, I., Deorte, R., Marynen, P., Gordts, S. and Cassiman, J.J. (1993) Co-amplification of the cystic fibrosis {Delta}F508 mutation with the HLA DQA1 sequence in single cell PCR: implications for improved assessment of polar bodies and blastomeres in preimplantation diagnosis. Prenat. Diagn., 13, 1111–1122.[ISI][Medline]

Submitted on June 11, 2003; resubmitted on August 5, 2003. accepted on August 11, 2003


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?


This article has been cited by other articles:


Home page
Nucleic Acids ResHome page
J. J. Li, Y. Chu, B. Y.-H. Lee, and X. S. Xie
Enzymatic signal amplification of molecular beacons for sensitive DNA detection
Nucleic Acids Res., April 1, 2008; 36(6): e36 - e36.
[Abstract] [Full Text] [PDF]


Home page
Proc. Natl. Acad. Sci. USAHome page
K. E. Pierce, J. A. Sanchez, J. E. Rice, and L. J. Wangh
Linear-After-The-Exponential (LATE)-PCR: Primer design criteria for high yields of specific single-stranded DNA and improved real-time detection
PNAS, June 14, 2005; 102(24): 8609 - 8614.
[Abstract] [Full Text] [PDF]


This Article
Right arrow Abstract Freely available
Right arrow FREE Full Text (PDF) Freely available
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrow Search for citing articles in:
ISI Web of Science (7)
Right arrowRequest Permissions
Google Scholar
Right arrow Articles by Pierce, K. E.
Right arrow Articles by Wangh, L. J.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Pierce, K. E.
Right arrow Articles by Wangh, L. J.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?